Abstract

Non-small cell lung cancer (NSCLC) is the most common type of lung cancer with substantial morbidity and mortality. Herein, a new signal-on electrochemiluminescence (ECL) immunosensor based on multiple amplification strategies is constructed for ultrasensitive detection of cytokeratin 19 fragment antigen 21-1 (CYFRA21-1) biomarker related to NSCLC. Polyethyleneimine (PEI) functionalized MXene is decorated with NiMn layer double hydroxide (NiMn LDH) to form MXene-PEI-NiMn LDH composite. Specially, the La-MOF@ZIF-67 bimetallic organic framework (named as LZBM) and MXene-PEI-NiMn LDH both served as coreaction accelerators to improve the ECL emission of the luminol-H2 O2 system. To be specific, Au nanoparticles (AuNPs) coated MXene-PEI-NiMn LDH is applied to immobilize primary CYFRA21-1 antibody (Ab1 ), while AuNPs decorated LZBM was used for the loading of luminol and secondary CYFRA21-1 antibody (Ab2 ) to form tracer label. Therefore, the ECL signal of the sandwich-type immunosensor is significantly enhanced due to the high loading capability for luminol and the synergistic catalytic ability for the decomposition of H2 O2 into reactive oxygen species (ROS). Under the optimal experimental conditions, the ECL immunosensor exhibited good analytical performances for CYFRA21-1 detection with a wide linear range (100 fg mL-1 -100 ng mL-1 ) and a low limit of detection (85.20 fg mL-1 ), providing a promising method for early diagnosis of NSCLC.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.