Abstract

To clarify the pathophysiological role of dynamic arterial properties in cardiovascular diseases, we attempted to develop a new control system that imposes desired aortic impedance on in situ rat left ventricle. In 38 anesthetized open-chest rats, ascending aortic pressure and flow waveforms were continuously sampled (1,000 Hz). Desired flow waveforms were calculated from measured aortic pressure waveforms and target impedance. To minimize the difference between measured and desired aortic flow waveforms, the computer generated commands to the servo-pump, connected to a side branch of the aorta. By iterating the process, we could successfully control aortic impedance in such a way as to manipulate compliance and characteristic impedance between 60 and 160% of their respective native values. The error between desired and measured aortic flow waveforms was 70 +/- 34 microl/s (root mean square; 4.4 +/- 1.4% of peak flow), indicating reasonable accuracy in controlling aortic impedance. This system enables us to examine the importance of dynamic arterial properties independently of other hemodynamic and neurohumoral factors in physiological and clinical settings.