Abstract

We propose a new, paper-based analytical device (PAD) for blood typing that allows for the simultaneous determination of ABO and Rh blood groups on the same device. The device was successfully fabricated by using a combination of wax printing and wax dipping methods. A 1:2 blood dilution was used for forward grouping, whereas whole blood could be used for reverse grouping. A 30% cell suspension of A-cells or B-cells was used for haemagglutination on the reverse grouping side. The total assay time was 10 min. The ratio between the distance of red blood cell movement and plasma separation is the criterion for agglutination and indicates the presence of the corresponding antigen or antibody. The proposed PAD has excellent reproducibility in that the same blood groups, namely A, AB, and O, were reported by using different PADs that were fabricated on the same day (n=10). The accuracy for detecting blood group A (n=12), B (n=13), AB (n=9), O (n=14), and Rh (n=48) typing were 92%, 85%, 89%, 93%, and 96%, respectively, in comparison with the conventional slide test method. The haematocrit of the sample affects the accuracy of the results, and appropriate dilution is suggested before typing. In conclusion, this study proposes a novel method that is straightforward, time-saving, and inexpensive for the simultaneous determination of ABO and Rh blood groups, which is promising for use in developing countries.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.