Abstract

Clot analogs are essential in animal and in vitro experiments on mechanical thrombectomy devices for treating acute ischemic stroke. Clot analogs should be capable of reproducing a variety of arterial clots observed in clinical practice in terms of histological composition and mechanical properties. Bovine blood with added thrombin was stirred in a beaker so that clots could be formed under the condition of dynamic vortical flow. Static clots were also prepared without stirring, and the properties of the static clots and dynamic clots were compared. Histological and scanning electron microscopy experiments were performed. Compression and relaxation tests were performed to evaluate the mechanical properties of the two types of clots. Thromboembolism and thrombectomy tests were conducted in an in vitro circulation model. Compared to the static clots, the dynamic clots prepared under vortical flow displayed a higher fibrin content, and their fibrin network was denser and sturdier than that of the static clots. The stiffness of the dynamic clots was significantly higher than that of the static clots. The stress of both types of clots could decay quickly under large sustained strain. The static clots could break at the bifurcation in the vascular model, while the dynamic clots could be firmly stuck in the vascular model. Dynamic clots generated in dynamic vortical flow differ significantly from static clots in terms of their composition and mechanical properties, which may be beneficial information for preclinical research on mechanical thrombectomy devices.