Abstract

Press-fit acetabular shells are subject to compressive loading conditions upon insertion during surgery. These conditions may result in deformation of the shell, potentially adversely affecting the performance of any liner fitted within it. This technical note describes the method used to measure acetabular shell deformation in a cadaver study using an optical GOM ATOS Triple Scan system. Six custom-made titanium acetabular shells were scanned prior to and immediately following insertion into three cadavers (one shell per acetabulum). A three-dimensional colour map was created for each shell, and cross sections were taken at the rim to determine the maximum diametrical deformation. Values of between 30 and 150 µm were recorded, with the results suggesting a size-dependent response.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.