Abstract
A novel method for surface temperature measurement using temperature sensitive paint (TSP) is presented. Precalibration of the TSP is shown and a semi in situ calibration technique using thermocouples is provided for high accuracy measurement. The method presented is applied to a film cooling experiment with a maximum surface temperature of 430 K and compared to highly reliable infrared thermography measurements that serve as benchmark results. The in situ calibration technique shows a maximum deviation of 0.5 K from the thermocouple readings. The comparison of laterally averaged temperature distributions of TSP and infrared measurement shows excellent agreement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.