Abstract

Evanescent wave (EW) broadband absorption spectroscopy is commonly interfaced with a range of analytical systems, including microfluidic flow cells, for identification and quantitation of species. A miniaturised spectrometer integrated with a microfluidic flow cell is useful to permit on-site analysis of samples. This work reports a novel leaky waveguide grating (LWG) device, which is able to obtain an absorption spectrum of an analyte of interest (in this case, methylene blue) without an external spectrometer. At 600 nm, the spectral resolution and the minimum detectable absorption coefficient of the LWG device is 49.5 nm and 1.65 cm(-1), respectively. The LWG can be fabricated on birefringent substrates such as plastics because the performance of the device is independent of the polarisation of the excitation source. The LWG device was fabricated using microcontact printing and hence can be easily mass produced.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.