Abstract

In this study using a discordant, xenogeneic, transplant model we demonstrate the functionality and safety of the first stent-based bioartificial pancreas (BAP) device implanted endovascularly into an artery, harnessing the high oxygen content in blood to support islet viability. The device is a self-expanding nitinol stent that is coated with a bilayer of polytetrafluoroethylene that forms channels to hold islets embedded in a hydrogel. We completed a 1-month study in the nondiabetic swine model (N = 3) to test the safety of the device and to assess islet functionality after device recovery. The luminal diameter of the devices from 3 animals on day 0 and day 30 was 10.01 ± 0.408 mm and 10.05 ± 0.25 mm, respectively. The stimulation index of the control and endovascular BAP devices explanted at day 30 were 3.35 ± 0.97 and 4.83 ±1.20, respectively, and the islets stained positively for insulin and glucagon after 30 days in vivo. This pilot study shows that BAP implantation into a peripheral artery is safe and supports islet functionality over 30 days, providing the groundwork for future work assessing the in vivo function of the device in diabetic swine.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.