Abstract

Goal: To develop and validatea novel neonatal non-invasive respiratory support device prototype designed to operate in low-resource settings. Methods: The device integrates a blower-based ventilator and a portable oxygen concentrator. A novel control algorithm was designed to achieve the desired fraction of inspired oxygen (FiO2) while minimizing power consumption. The accuracy of the delivered FiO2 and the device power consumption were evaluated in vitro, and a formative usability test was conducted in a rural hospital in Uganda. Results: The agreement between the set and delivered FiO2 was high (limit of agreement:-5.6 ÷ 3.8%). For FiO2 below 60%, the control algorithm reduced the power drain by 50%. The device was also appreciated by intended users. Conclusion: The prototype proved effective in delivering oxygen-enriched continuous positive airway pressure in the absence of compressed air and oxygen, holding promise for a sustainable and effective implementation of neonatal respiratory support in low-resource settings.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.