Abstract
AbstractIn this study, a molecularly imprinted sensor technology is engineered to detect glucose in real blood samples by chronoimpedimetrically. The imprinting process of glucose (Glc) was carried out by electrochemical polymerization of aminophenylboronic acid (APBA) and pyrrole (Py) by performing cyclic voltammetry (CV). Afterwards, glucose molecule was removed from imprinted surface by 5 % acetic acid to reveal glucose imprinted cavities. Electrochemical Impedance Spectroscopy (EIS) was used to characterize sensor modification steps and glucose removal. Glucose monitoring process was carried out chronoimpedimetrically(CI) for the first time in real blood samples. Calibration curve was prepared between 20–800 mg/dL. The standard deviations of the 18 calibration curves R2 were calculated as 0.9866±0.0066 to assess reproducibility. Recovery was calculated by using 105 mg/dL Glc Serum Sample, which was monitored by auto analyzer and into this sample 50 mg/dL Glc added and our sensor response was 147.92±2.43 mg/dL, 98.6±1.62 % (n=5). Non‐imprinted (NIP) sensor gave no signal for the glucose concentration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.