Abstract

A system is described which introduces a cooling system based on a proportional liquid nitrogen control valve. This is driven directly from the 3 term P.I.D. output of the DSC controller which in turn is controlled directly from the Polymer Laboratories Limited software system. The proportioning valve adjusts the amount of liquid nitrogen required by the DSC as determined by the P.I.D. control algorithm. By varying the nitrogen flow according to demand, absolute repeatability is ensured from run to run and a considerable saving in liquid nitrogen is achieved. Thus the DSC performs identically in heating or cooling mode.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.