Abstract

The purpose of this study was to evaluate the usefulness of the injection pressure-to-injection rate (IPIR) ratio for the early detection of contrast extravasation at the venipuncture site during contrast-enhanced computed tomography. We retrospectively enrolled 57,528 patients who underwent contrast-enhanced computed tomography examinations in a single hospital. The power injector recorded the contrast injection pressure at 0.25-second intervals. We constructed logistic regression models using the IPIR ratio as the independent variable and extravasation occurrence as the dependent variable (IPIR ratio models) at 1, 2, 3, 4, 5, and 6 seconds after the start of contrast administration. Univariate logistic regression models in which injection pressure is used as an independent variable (injection pressure models) were also constructed as a reference baseline. The performance of the models was evaluated with the area under the receiver operating characteristic curves. Of the 57,528 cases, 46,022 were assigned to the training group and 11,506 were assigned to the test group, which included 112 extravasation cases (0.24%) in the training group and 28 (0.24%) in the test group. The area under the receiver operating characteristic curves for the IPIR ratio models and injection pressure models were 0.555 versus 0.563 at t = 1 (P = 0.270), 0.712 versus 0.678 at t = 2 (P = 0.305), 0.758 versus 0.693 at t = 3 (P = 0.032), 0.776 versus 0.688 at t = 4 (P = 0.005), 0.810 versus 0.699 at t = 5 (P = 0.002), and 0.811 versus 0.706 at t = 6 (P = 0.002). The IPIR ratio models perform better in detecting contrast extravasation at 3 to 6 seconds after the start of contrast administration than injection pressure models.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call