Abstract

PEGylation has been used as a strategy to enhance pharmacokinetic properties of therapeutic proteins by pharmaceutical industry. Imaged CIEF (iCIEF) is the current industry standard technology for pI determination and charge variant quantification of proteins and antibodies. However, the charge variants of PEGylated proteins merge into one broad peak during iCIEF, most likely due to masking of proteins by the surrounding PEG chain as well as the increased hydrodynamic volume due to PEGylation. Here, we report our novel matrix formula with a combination of glycine and taurine that significantly improved the separation of charge variants in PEGylated proteins. As a result, it is no longer necessary to conduct IEF of proteins prior to PEGylation, which does not reflect the changes caused by PEGylation and purification processes. The novel matrix (glycine and taurine) enables iCIEF analysis of PEGylated proteins in their real conjugated states.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.