Abstract

Simultaneous determination of aerodynamic particle size distribution (APSD) of drug particles of mometasone furoate and formoterol fumarate dihydrate – 200 mcg/5 mcg inhalation aerosol (MDI) has been developed and reported in this work. The method was developed in consistent with USP <601> inhalation and nasal drug products e.g. aerosols, sprays and powders. The next generation impactor (NGI) was used and operated at 30 ± 5% L/min. The proposed method was demonstrated for sensitivity, minimum number of actuations, diluent volume optimization, mass balance and precision. The proposed method was found to be robust, accurate, linear and precise.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call