Abstract

ObjectivesTo evaluate the feasibility and safety of a new unibody branched stent-graft for reconstruction of the canine aortic arch. Materials and methodsTwenty adult hybrid dogs were used for the experiments. Ten dogs were implanted with single-branched stent-grafts; the other ten dogs were implanted with double-branched stent-grafts. The stent-grafts were implanted transluminally via the abdominal aorta. The branched limbs were caught and pulled into supra-aortic vessels using gooseneck snare wires introduced via the axillary arteries. The animals were euthanized 4 months after implantation. ResultsOne of the ten dogs implanted with a single-branched stent-graft died from failure of the implantation procedure, and two of the ten dogs implanted with double-branched stent-grafts died from failure of the procedure and excessive blood loss. After month 4, the remaining unibody branched stent-grafts were patent and did not migrate. ConclusionsThis new unibody branched stent-graft could be used to reconstruct the aortic arch. This is a total endovascular technique, and compared to other branched stent-grafts appears to be safer and easier to implant.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.