Abstract
This article introduces a new tool to measure the pressure that is under pressure garments. The Iscan (Tekscan, Inc.) system uses a patented ultra-thin (0.007 inch) sensor with multiple sensing locations that sample continuously at 100 times per second. It is noninvasive, convenient, and quick. The study had two parts. First, we established the validity and reliability of the device. Next, garment/scar interface pressures were measured on new garments with use of the Iscan system. Four garment types were studied, with 10 measurements made in each group: Isotoner gloves (Smith & Nephew Roylan, Inc.); custom-fit pressure gloves; Tubigrip forearm sleeves (Seton Health Care Group); and custom-fit pressure forearm sleeves. Mean garment/scar interface pressures were 18 +/- 2 mm Hg for the Isotoner glove, 34 +/- 5 mm Hg for the custom-fit pressure glove, 20 +/- 7 mm Hg for the Tubigrip sleeve, and 35 +/- 6 mm Hg for the custom-fit sleeve. We concluded that the Iscan system can be used to measure pressure under pressure garments accurately and reliably, and that custom-fit hand and forearm garments provide more pressure than Isotoner gloves or Tubigrip sleeves.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.