Abstract
A new technique was developed for measuring the concentrations of airborne radon progeny by using an imaging plate (IP). The concentrations of radon progeny collected on the sampling filter were measured with the IP at 2-5-, 6-20- and 21-30-min intervals after sampling, and the alpha disintegrations were automatically counted by a self-developed image analysis software. To reduce the overlapping impact on alpha counts, a plastic separator with a 0.5-mm thickness was set between the sampling filter and the IP. The lower limit of detection for the equilibrium equivalent concentration of radon was estimated to be ∼3.5 Bq m(-3) for a 10-min sampling at 10 l min(-1). Comparison experiments showed that results measured with the new technique were well consistent with those measured with the alpha spectrometer. It indicates that the new technique is a useful method for accurate measurements of airborne radon progeny concentrations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.