Abstract
In this paper a new quality indicator for two-tier calibration procedures that use only reflection standards is presented and applied to coaxial-to-waveguide transitions. The quality indicator is based on the algebraic conditioning of the system of equations solved for obtaining transition characteristics. The study has been carried out in a wide bandwidth as a difference with previous works. The obtained results indicate that a threshold value for this indicator around 10% can be established. For values below this limit the error grows to unacceptable values. Additionally, it has been shown that an exponential relationship between quality indicator and the error can be predicted.
Paper version not known (Free)
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.