Abstract

In the current study, a novel double-layer gold nanoparticles- electrochemical immunosensor electrode (DGN-EIE) immobilized with Salmonella plasmid virulence C (SpvC) antibody was developed. To increase the fixed quantity of antibodies and electrochemical signal, an electrochemical biosensing signal amplification system was utilized with gold nanoparticles-thionine-chitosan absorbing horseradish peroxidase (HRP). In addition, the SpvC monoclonal antibodies (derived from Balb/c mice) were prepared and screened with a high affinity to SpvC. To evaluate the quality of DGN-EIE, the amperometric I-t curve method was applied to determine Salmonella in PBS. The results showed that the response current had a good linear correlation with the bacterial quantity ranged from 1.0 × 101-5.0 × 104cfu/mL. The lowest detection limit was found at 5cfu/mL. Furthermore, the proposed immunosensor has been demonstrated with high sensitivity, good selectivity and reproducibility. Apparently, DGN-EIE may be a very useful tool for monitoring the bacteria.