Abstract
The enthalpies of sublimation of naphthalene, benzoic acid and phenanthrene have been determined by isothermal operation of a DSC, coupled to a vacuum system. The results obtained by this method are in excellent agreement with those reported in the literature from established procedures. The precision obtained was found to be dependent on the rate of sublimation. This paper describes the experimental procedure, calibration, the corrections applied and comparison of the results, mainly with those obtained by dynamic experiments in DSC.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.