Abstract
Abstract A new materials irradiation facility with improved control capabilities has been installed at the Kyoto University Reactor (KUR). Several deficiencies of conventional fission neutron material irradiation systems have been corrected. The specimen temperature is controlled both by an electric heater and by the helium pressure in the irradiation tube without exposure to neutrons at temperatures different from the design test conditions. The neutron spectrum is varied by the irradiation position. Irradiation dose is changed by pulling the irradiation capsule up and down during irradiation. Several characteristics of the irradiation field were measured. The typical irradiation intensity is 9.4×10 12 n/cm 2 s (>0.1 MeV) and the irradiation temperature of specimens is controllable from 363 to 773 K with a precision of ±2 K.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.