Abstract

This paper describes how to calibrate the right angle phase of a Four-Terminal-Pair (4TP) LCR meter. The meter uses the right angle signal to process synchronous detector as a vector receiver. The right angle calibration method uses, at least, one programmable phase signal generator and a phase shift circuits. The signal generator and the phase shifter do not require a precision 90 degree angle. As a result, the right angle phase deviation of the LCR meter is better than 100 murad at 10 MHz. This method is also applicable to the phase calibration of a vector network analyzer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.