Abstract

This paper describes a new inductance measurement system based on a commercially available LCR meter, a commercially available audio frequency capacitance bridge, 4 terminal-pair (4TP) capacitance and resistance standards, and a multi-stage, programmable-ratio, current transformer. The described measurement system synthesizes equivalent inductance values using the current transformer to combine the applied currents through programmable resistance and capacitance standards. The system takes advantage of the inherent measurement linearity of the LCR meter, and achieves acceptable inductance measurement uncertainty over the 50 Hz to 20 kHz frequency range

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.