Abstract
We describe a new high performance liquid chromatography coupled with ultraviolet detection method for the quantification of plasma concentration of oral iron chelating agent deferasirox. A simple protein precipitation extraction procedure was applied on 500μl of plasma aliquots. Chromatographic separation was achieved on a C18 reverse phase column and eluate was monitored at 295nm, with 8min of analytical run. This method has been validated following Food and Drug Administration procedures: mean intra and inter day variability was 4.64 and 10.55%; mean accuracy was 6.27%; mean extraction recovery 91.66%. Calibration curves ranged from 0.078125 to 40μg/ml. Limit of quantification was set at 0.15625 while limit of detection at 0.078125μg/ml. We applied methodology developed on plasma samples of thalassaemic patients treated with deferasirox, finding correlation between deferasirox plasma concentrations and serum ferritin levels. This methodology allowed a specific, sensitive and reliable determination of deferasirox, that could be useful to perform its therapeutic monitoring and pharmacokinetic studies in patients plasma.