Abstract
A novel polymeric membrane sensor was developed by using 2-hydroxyethyl acrylate, trimethylolpropane triacrylate, (3-mercaptopropyl)trimethoxysilane, and poly(ethylene glycol)diacrylate for As(III) determination. Various parameters, like the pH, response time, the foreign ions, and concentration effects were investigated for deciding the optimum working conditions of the polymeric sensor. As a result of this investigation, the optimum pH was found to be 2, and the response time was found to be 30 s. The linear range of the sensor was 6.65 × 10-9 - 3.99 × 10-8 mol L-1 (0.50 - 2.99 μg L-1) with a detection limit of 2.33 × 10-9 mol L-1 (0.18 μg L-1). Soy flour and well-water samples were successfully analyzed with the developed sensor. The sensor can be used at least 100 times after regeneration. It could be reused by washing with purified water and showed good stability for 6 months.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Analytical sciences : the international journal of the Japan Society for Analytical Chemistry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.