Abstract

We describe a Reference and Switching Unit (RSU), that, with a digital multimeter (DMM) and a dc voltage calibrator, constitutes a system for calibration of standard resistors in the range 100 kΩ–10 TΩ at measurement voltages up to 1000 V. This system represents an implementation of a DMM-based system already developed at IEN. We report the characteristics of the DMM-based system, the metrological and electronic features of the RSU, the uncertainty levels that can be achieved with the DMM-based and whit the RSU-based systems. Finally, we show the obtained results in order to check the compatibility of the two systems in the calibration of standard resistors in the range 1 GΩ–1 TΩ. The 2 σ best measurement capabilities of the RSU-based system span from 6.3×10 −6 for the calibration of a 100 kΩ standard resistor at 10 V, to 1.2×10 −3 for the calibration of 10 TΩ standard resistor at 1000 V.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.