Abstract
Medical imaging has long benefited from advances in photon counting detectors arising from space and particle physics. We describe a microchannel plate-based detector system for high content (multi-parametric) analysis, specifically designed to provide a step change in performance and throughput for measurements in imaged live cells and tissue for the ‘omics’. The detector system integrates multi-channel, high time resolution, photon counting capability into a single miniaturized detector with integrated ASIC electronics, comprising a fast, low power amplifier discriminator and TDC for every channel of the discrete pixel electronic readout, and achieving a pixel density improvement of order two magnitudes compared with current comparable devices. The device combines high performance, easy reconfigurability, and economy within a compact footprint. We present simulations and preliminary measurements in the context of our ultimate goals of 20 ps time resolution with multi-channel parallel analysis (1024 channels).
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have