Abstract

In this work, we report an energy-efficient switched capacitor based millimeter-scale pacemaker (5 mm ×7.5 mm) and a multi-receiver wireless energy transfer system operating at around 200 MHz, and use them in a proof-of-concept multi-site heart pacing study. Two pacemakers were placed on two beating Langendorff rodent heart models separately. By utilizing a single transmitter positioned 20-30 cm away, both Langendorff hearts captured the stimuli simultaneously and were electromechanically coupled. This study provides an insight for future energy-efficient and distributed cardiac pacemakers that can offer cardiac resynchronization therapies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.