Abstract

In order to set up magnetic resonance imaging (MRI) procedures of arbitrary voxel dimensions, slice orientation, and sequence timing in a reasonable time, some form of automatic gradient pulse calibration is required. One such method, involving simulation of gradient waveforms, is presented. Waveforms are modeled based on measurements of the step response. The model used divides each transition into three time regions: a "start" region in the first 0.3 ms, a "slew" region, and a "tail" region representing decay of the eddy current compensation error. In the "slew" region, the time derivative of the gradient, G' (t), is expressed as a function of G(t). The first two regions are nonlinear with respect to demand. The mean error in the simulated gradient is generally less than 0.04 mT m-1 in spin echo sequences. Image signal/noise ratios resulting from sequences calibrated using the model are within 5% of those of empirically calibrated sequences.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.