Abstract
In paediatric radiography, according to the as low as reasonably achievable (ALARA) principle, the imaging task should be performed with the lowest possible radiation dose. This paper describes a Monte-Carlo simulation framework for dose optimisation of imaging parameters in digital paediatric radiography. Patient models with high spatial resolution and organ segmentation enable the simultaneous evaluation of image quality and patient dose on the same simulated radiographic examination. The accuracy of the image simulation is analysed by comparing simulated and acquired images of technical phantoms. As a first application example, the framework is applied to optimise tube voltage and pre-filtration in newborn chest radiography. At equal patient dose, the highest CNR is obtained with low-kV settings in combination with copper filtration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.