Abstract

A modified spin-echo pulse sequence is described that enables accurate T 2 measurements to be made in NMR microimaging experiments. The modified sequence eliminates cumulative diffusion losses that lead to an underestimation of the T 2 relaxation time using conventional spin-echo pulse sequences. The approach is theoretically justified and confirmed in comparative experiments on phantoms.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.