Abstract

For the Clinac 4, open field profiles measured in the principal plane have higher intensity "horns" than those in off-axis planes. The maximum deviation occurs at 1-cm depth, where in the worst case of large field sizes and off-axis distances the deviation can be as high as 16% in the region near the horns. A model is proposed to generate open field beam profiles in off-axis planes, based on measured profiles in the transverse principal plane and in the largest field size diagonal plane of the machine. Within the central 90% portion of the field, the maximum deviation of the generated profiles from the measured ones at the same off-axis distance varies from about 3% at 1-cm depth to about 2% at 13-cm depth and then increases to less than 5% at 25-cm depth, even for very large field sizes and off-axis distances. Very little additional computer time and data storage are required for this procedure. Using these profiles, the Milan and Bentley method can be extended to three-dimensional treatment planning with good accuracy.