Abstract
An ion trap mass spectrometer is conventionally featured with an electron multiplier as its detector. However, an electron multiplier can typically work at pressures below 20 mTorr with a high voltage applied, which limits the further miniaturization of ion trap mass spectrometers. In this work, a low noise Faraday detector was developed and integrated in our miniature mass spectrometer instrument, and a post data processing method was applied to improve its performance. A limit of detection of 1 ng mL-1 was achieved, and quantitation performance and mass resolution were characterized. This technology could be useful in the further development of miniature mass spectrometers by increasing background pressures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.