Abstract
A mild HPLC method for the quality control of [18F]sodium fluoride solution for injection was developed. The method used anion exchange column for separation, UV and online radio-detectors for fluoride content and fluorine-18 determination. A mixture of 100 mM sodium acetate and 25 mM sodium chloride was chosen as mobile phase. The method was validated for system suitability, specificity, fluoride concentration and radiochemical purity determination with good linearity, accuracy and precision. The good robustness of the system was also verified, enabling routine analysis of fluoride content and radiochemical purity of the [18F]sodium fluoride product.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.