A compact gas analyzer for monitoring respiratory gases in the operating room has been designed using a low resolution quadrupole mass filter. The vacuum is generated by a small diffusion pump and controlled by a microprocessor which also converts the ion currents from the quadropole into gas concentration values using sensitivity factors calculated during automatic calibration against a built-in five component gas mixture.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.