Abstract

The rheology of air or moisture sensitive liquids, gels, and glasses requires complicated rheometer-in-glovebox laboratory setups. Here, we demonstrate the use of a heavier-than-air cover gas, sulfur hexafluoride, and the design of a cover gas container that can attach to the lower geometry plate of any rheometer to carry out rheology experiments on air-sensitive liquids and soft solids. Rheological measurements of air-reactive ionic liquid trimethylsulfonium bromide-AlCl3, moisture sensitive titanium(IV) propoxide, and glycerin demonstrate the effectiveness of the cover-gas method for loading samples on acquiring correct temperature dependent viscosity data of the sample in the absence of reaction products.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call