Abstract

A diffusive gradients in thin films (DGT) device for the analysis of free Cd(II) species, based on Cd(II) ion-imprinted sorbent (IIS) as the binding agents and commercial polyethersulfone membrane (PES) as diffusion layer, was developed (PES/IIS-DGT). DGT time-series experiments showed that the mass of free Cd(II) species accumulated by PES/IIS-DGT was linear vs. time (R2 = 0.9953) and the concentration of free Cd(II) species by PES/IIS-DGT was in good agreement with the total dissolved concentrations of free Cd(II) species in simple synthetic solutions where free ionic species dominated. PES/IIS-DGT performance was independent in the range of pH 4.5–7.5 and ionic strength range from 1.0 × 10−3 to 0.7 mol L−1. The measurement of free Cd(II) species in synthetic solution containing different concentrations of ligands by PES/IIS-DGT showed an excellent agreement with the value measured by Cd(II) ion selective electrodes (Cd-ISE), indicating that PES/IIS-DGT method is more suitable than Cd-ISE for the measurement of low concentration of free Cd(II) species due to the enrichment of IIS for the analytes.