Abstract

This paper presents a low-power analog front-end that enables photoplethysmographic signals acquisition, the dynamic range for AC component exaction is relaxed with simple high-pass implementation. The chopping modulation ensures the low-noise operation. The circuit is fabricated in a 0.18-um CMOS technology. Measurements show that the consuming current is approximately 72 uA at a supply of 2.5 V. The circuit achieves a input noise of 6.45 $\mathrm{pA}/\sqrt{Hz}$. The calibred algorithm is implemented by means of MCU, and the demonstration that is compared with the Fluck Simulator used as the reference shows the heart rate is accurately detected, and the error of the measured blood oxygen saturation is less than 1.5%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.