Abstract

A small-mass system has been developed for monitoring the flux of neutrons with energy up to 1 MeV at the new time-of-flight facility at CERN, n_TOF. The monitor is based on a thin Mylar foil with a 6Li deposit, placed in the neutron beam, and an array of silicon detectors, placed outside the beam, for detecting the products of the 6Li( n, α) 3H reaction. The small amount of material on the beam ensures a minimal perturbation of the flux and minimizes the background related to scattered neutrons. Moreover, a further reduction of the γ-ray background has been obtained by constructing the scattering chamber hosting the device in carbon fibre. A detailed description of the flux monitor is here presented, together with the characteristics of the device, in terms of efficiency, resolution and induced background. The use of the monitor in the measurement of neutron capture cross-sections at n_TOF is discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.