Abstract
The performance of a new, low-background NaI(Tl) spectrometer, based on γγ-coincidence counting, is discussed. We present experimental coincidence efficiencies, timing resolutions, background count rates, and minimum detectable activities. The spectrometer has been simulated using Geant4, and the results are used for estimating coincidence efficiencies for volume sources. To test the device, we measured the cosmogenic 26Al activity in a small (17.7 g) meteorite fragment. We find a value of 52.9±7.8dpm∕kg, in agreement with the activity measured previously in a much larger fragment of the same meteorite using a HPGe detector.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.