1. He thought the minimum detectable activity (MDA) value for K is too low, the counting time was too short based on his experience and why distilled water was used. First of all the distilled water was used to limit the level of backscatter radiation within the detector system during the background count. Also, we want to state that a low background spectrometry system was used for the counting. In other to optimize the result of the background count and the real samples, the same counting time was used. We agree that the longer the counting time the greater the peak area, however, based on the prevailing laboratory condition which could not enable us to consider a longer counting time 36,000 was used which could also give an appreciable count rate. The MDA is defined as the smallest quantity of radioactivity that could be measured under specified conditions, and is another factor which is an important concept in environmental level measurement. The MDA depends on the lower limit of detection (LLD) and the counting efficiency of a counting system. The MDA is very important, particularly in environmental level systems, where the count rate of a sample is almost the same as the count rate of the background. The MDA was calculated using the conventional MDA method based on statistical convergence of 95 % confidence level. The value obtained for K was based on experimental data and calculation done using the conventional method.