Abstract
A rapid and sensitive liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed and validated for simultaneous quantification of escin Ia and escin Ib in human plasma. After a solid-phase extraction (SPE), the analytes were separated on a Zorbax Extend C(18) column by isocratic elution with a mobile phase of methanol-acetonitrile-10 mm ammonium acetate (27:27:46, v/v/v) at a flow rate of 1.0 mL/min and analyzed by mass spectrometry in the positive ion multiple reaction monitoring mode. The precursor to product ion transitions of m/z 1131.8 → 807.6 was used to quantify escin Ia and escin Ib. Good linearity was achieved over a wide range of 2.00-900 ng/mL for escin Ia and 1.50-662 ng/mL for escin Ib. The intra- and inter-day precisions (as relative standard deviation) were less than 11% for each QC level of escin Ia and escin Ib. The accuracies (as relative error) were within ±5.27% for escin Ia and within ±4.07% for escin Ib. The method was successfully employed in a pharmacokinetic study after a single intravenous infusion administration of sodium aescinate injection containing 10 mg escin to each of the 10 healthy volunteers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.