Abstract

A rapid and sensitive liquid chromatography-tandem mass spectrometry (LC/MS/MS) method to determine levonorgestrel in human plasma was developed and fully validated. After hexane-ethyl acetate (70:30, v/v) induced extraction from the plasma samples, levonorgestrel was subjected to LC/MS/MS analysis using electro-spray ionization. The MS system was operated in the selected reaction monitoring mode. Chromatographic separation was performed on a Hypersil BDS C18 column (i.d. 2.1x50 mm, particle size 3 microm). The method had a chromatographic running time of 2.0 min and linear calibration curves over the concentration ranges of 0.25-90 ng/mL for levonorgestrel. The lower limit of quantification of the method was 0.25 ng/mL for levonorgestrel. The intra- and inter-batch precision was 3.7-10.2 and 5.1-12.9%, respectively, for all quality control samples at concentrations of 0.5, 6.0 and 45.0 ng/mL. These results indicate that the method was efficient with a simple preparation procedure and a very short running time (2.0 min) for levonorgestrel compared with those methods reported in the literature and had high selectivity, acceptable accuracy, precision and sensitivity. The validated LC/MS/MS method was successfully used for a bioequivalence study of two tablet formulations of levonorgestrel in healthy volunteers.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.