Abstract

Laser desorption/ionization (LDI) mass spectrometry suffers from poor reproducibility and a strong bias in representing the chemical composition of the sample in the mass spectra. In regard to develop laser mass spectrometry towards a quantitative analysis technique desorption and ionization processes should be separated. Here we describe a time-of-flight mass spectrometer (TOF-MS) with laser ablation and subsequent electron impact ionization (LA-EI-MS). Sample material is ablated by an Nd-YAG laser pulse (λ=355 nm, E=0.5–35 mJ, τ=9 ns), which produces ions and neutrals in the ablation plume. After expansion, parts of the plume enter the EI ion source where neutrals are ionized by electron impact (Eel=70 eV). Ions are extracted from the ion source into the flight tube of the TOF-MS by applying a pulsed repeller potential. Depending on delay times between laser pulse and repeller pulse either neutrals or ions emitted from the sample were analyzed. Different mass spectra of the same sample are obtained in these two analyzing modes under the same irradiation conditions. With a relative standard deviation of 10% to 15%, reproducibility of mass spectra is significantly better in LA-EI-MS than in LDI-MS.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.