Abstract

A design for a radio frequency (RF) neutron spin flipper obtained from magneto-static and neutron spin transport simulations is presented. The RF flipper constructed from this design provides a flipping probability of 0.999 or better for a beam size 6cm wide and 15cm high and a wavelength band between 0.4 and 0.6nm. Three permanent magnet guide field sections with air gaps provide a linear field gradient along the beam propagation direction over a large cross-sectional area. An RF oscillator based on coupling the resonant coil of a Hartley oscillator to the excitation coil was developed, which provides a higher current and, thereby, a larger RF amplitude, as compared to a conventional RF power amplifier. Two opaque He3 neutron spin filters were employed to measure the flipping probability of the flipper with very high precision. A spatially uniform flipping probability of 0.9995(2) or higher was measured over the large cross-sectional area neutron guide. This RF neutron spin flipper will be employed in a polychromatic beam reflectometer at the National Institute of Standards and Technology Center for Neutron Research. This design can be applied to other polarized neutron instruments or applications requiring a very high continuous flipping probability of the neutron spin for a large cross-sectional area beam.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call