Abstract
This report introduces an on-chip integrated reverse transcription (RT)-microchip, which includes two genetic functionalities of RNA extraction and cDNA synthesis. In the RNA extraction compartment, RNA is extracted from peripheral blood lysate within 1min, by lateral magnetophoresis using magnetic oligo-dT beads. The extracted RNA is then collected and used directly to produce cDNA in the cDNA synthesis microchamber, which is monolithically integrated with the RNA extraction compartment. To verify the superiority of the proposed RT-microchip, RT-PCR amplification was performed using cDNA harvested from the RT-microchip, and the results were compared with those obtained using typical RNA extraction methods such as a silica matrix column and magnetic oligo-dT beads. The RT-PCR amplification results using 100μl of blood showed that the intensity of the bands in gel electrophoresis of the RT-microchip was 2-fold stronger than that of the silica matrix column and 2.65-fold stronger than that of the magnetic oligo-dT beads. The results demonstrate that the RT-microchip technique is the most sensitive of the tested methods.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.