Abstract A first generation amperometric glucose biosensor based on Prussian Blue modified electrodes was developed. Besides exception of noble metals (platinum as usual) the developed biosensor possessed high sensitivity. Linear response dependence on analyte concentration was observed in a range of 10−6 − 5·10−3 M glucose. The current density produced by addition of 10−6 M glucose was 0.18 μ A/cm2. Hydrogen peroxide produced via enzyme reaction was detected by electroreduction. Owing to that reason the biosensor response became independent of the presence of reductants (ascorbate for example). The developed amperometric biosensor was expected to obey requirements for non-invasive diagnostics. There are not principle limits of using other oxidases for Prussian Blue based amperometric biosensor development. Thus one can develop such biosensors for cholesterol, alcohol, glycerol, amino acids ets.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.