Abstract

A highly-sensitive glucose biosensor amenable to ultra-miniaturisation was fabricated by immobilisation of glucose oxidase (GOx), onto a poly(2,6-diaminopyridine)/multi-walled carbon nanotube/glassy carbon electrode (poly(2,6-DP)/MWNT/GCE). Cyclic voltammetry was used for both the electrochemical synthesis of poly-(2,6-DP) on the surface of a MWNT-modified GC electrode, and characterisation of the polymers deposited on the GC electrode. The synergistic effect of the high active surface area of both the conducting polymer, i.e., poly-(2,6-DP) and MWNT gave rise to a remarkable improvement in the electrocatalytic properties of the biosensor. The transfer coefficient (α), heterogeneous electron transfer rate constant and Michaelis–Menten constant were calculated to be 0.6, 4s−1 and 0.20mM at pH 7.4, respectively. The GOx/poly(2,6-DP)/MWNT/GC bioelectrode exhibited two linear responses to glucose in the concentration ranging from 0.42μM to 8.0mM with a correlation coefficient of 0.95, sensitivity of 52.0μAmM−1cm−2, repeatability of 1.6% and long-term stability, which could make it a promising bioelectrode for precise detection of glucose in the biological samples.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call