Abstract
A high-resolution LEED system with optimum operating parameters for use below 100 eV has been designed and tested. A field-emission-based electron gun with an angular divergence of 2 mrad, and a large sample detector distance of 150 mm provide the major improvements in resolution. Measured beam widths (Δk/2π FWHM) for both specular and nonspecular beams are approximately 0.002 Å−1 in the energy range of 50–100 eV.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.