Abstract

We present the design and performance of a high-pressure scanning tunneling microscope (HP–STM), which allows atom-resolved imaging of metal surfaces at pressures ranging from ultrahigh vacuum (UHV) to atmospheric pressures (1×10−10–1000 mbar) on a routine basis. The HP–STM is integrated in a gold-plated high-pressure cell with a volume of only ∼0.5 l, which is attached directly to an UHV preparation/analysis chamber. The latter facilitates quick sample transfer between the UHV chamber and the high-pressure cell, and allows for in situ chemical and structural analysis by a number of analytical UHV techniques incorporated in the UHV chamber. Reactant gases are admitted to the high-pressure cell via a dedicated gas handling system, which includes several stages of gas purification. The use of ultrapure gasses is essential when working at high pressures in order to achieve well-defined experimental conditions. The latter is demonstrated in the case of H/Cu(110) at atmospheric H2 pressures where impurity-related structures were observed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.